ASCII Programming
MI3 Rev. G Nov/2015 127
18.8.1 ASCII Commands for Ethernet and Profinet
Description
Char
Format
P
B
S
N
Legal Values
Factory default
Head /
Box
IP address
IP
192.xxx.xxx.xxx
192.168.42.130
(Ethernet)
0.0.0.0
(Profinet)
B
2.19
Net Mask
NM
255.255.255.0
255.255.255.0
(Ethernet)
0.0.0.0
(Profinet)
B
2.19
Gateway
GW
192.xxx.xxx.xxx
192.x.x.x
(Ethernet)
0.0.0.0
(Profinet)
B
2.19
MAC address
(read only)
MAC
001d8d 1xxxxx
00:1d:8d:10:00:14
(Ethernet)
00:1d:8d:10:00:1f
(Profinet)
B
2.19
PORT address
(Ethernet only)
PORT
0 65535 (TCP/IP only)
6363
B
2.19
DHCP on/off
(Ethernet only)
IPU
0 off
1 - on
0
B
2.19
Data logging on/off
(Ethernet only)
DL
0 off
1 - on
0
B
2.19
Data logging interval
(Ethernet only)
DLI
1 2097120 s
1s
B
2.19
Get Ethernet FW
Version
ETV
String, 4 char
1.00
B
2.19
Handle Profinet Name
PNN
String, 10 char
commbox
B
2.19
Reset Ethernet module
RSE
B
2.19
TCP/IP time out interval
TTI
TTI=0
1…240 s
0 … infinitely
120
B
2.20
Web Server on/off
1
WS
?WS
0 - off, 1 - on
1
B
2.20
Table 18: Command Set
1
WS is OFF while Burst mode is ON and Ethernet communication is running