8 Cleaning Configuration
The Real Clean system is controlled and configured through the M series analyzer. Below is a general
summary of the functions and configurations for the system.
Table 4: Menu functions and configuration for Real Clean System
MENU
DESCRIPTION
-Cleaning System
Enable Clean
Cleaning must be enabled for the Real Clean system to operate.
Reset Clean
When the cleaning fluid is changed or refilled, the user must do a Reset to reset the counter on
the cleaning bottle.
Force Clean
At any time, the user can trigger a cleaning cycle by selecting force clean. This does not
interfere with the normal cleaning schedule.
Pumping Time
Pumping Time allows for configuration of time (in seconds) the cleaning system pumps
cleaning fluid to the sensor. The pumping time should be configured to ensure the cleaning
fluid reaches and fills the quartz flow cell prior to the 10-minute holding period. Longer
distances between the cleaning system and sensor will require a longer pumping time. Default
setting is 120 seconds, suitable for a tube length of 24-36 inches.
-Clean Frequency
Clean Frequency allows the frequency of cleaning to be configured i.e. hour, day, week, or
month. Default is 3 days.
-Alarms
If enabled, alarm conditions are displayed on the main menu.