Real Clean System I v 2.1
Real Tech Inc. 4
2 Overview
Thank you for purchasing from Real Tech Inc.
The Real Clean System I is an automatic chemical cleaning system for use with
any Real Tech Inc. M2000 or M3000 series analyzers. The purpose of the
system is to reduce fouling of the flow cell within the analyzer.
The Real Clean and M series Analyzer are plumbed so that process water goes through the Real Clean to the
Analyzer. When a cleaning cycle starts, the Real Clean system halts input of process water and pumps
cleaning solution from its wash bottle to the flow cell of the M series analyzer on a timed schedule, or when
needed, immediately through manual intervention.
A portion of cleaning fluid is recaptured and directed back into the wash bottle after a cleaning cycle is
completed. Analyzer measurement resumes after the tubing is cleared of residual cleaning fluid by influx of
process water.
The cleaning system is controlled through the Operator Interface of the Real Tech M series analyzer. From
here, the system is enabled, and a cleaning frequency is selected.
Cleaning of the flow cell is best achieved when cleaning frequency and cleaning fluid work in unison to prevent
the accumulation of fouling agents inside the flow cell.
*Your Serial Number is located on the inside of the cabinet door.