Register functions carried out in Supervisor Mode are as follows:
• Program prices, taxes and discounts for up to 10 individually named products
• Set the time and date for the E4000, printed on delivery, end-of-shift and calibration reports.
• Program the register ow control parameters
• With no valve control (E4000 relays do not operate a valve)
• With non-preset valve control (LPG security solenoid valve)
• With preset valve control (2 stage shut off valve)
• Congure the communication ports for the printer and mobile computer connection
• Allow multiple deliveries (disable 3 minute timeout to print ticket)
• Edit delivery ticket number, supervisor password, truck number
• Program a miscellaneous fee
Note: Complete software details are listed in the E4000 Software Manual
5.1.3 W&M Mode (Weights and Measures Mode)
The W&M Mode is the menu selection allowing a technician to calibrate the E4000 to the owmeter it
will operate with. The Weights and Measures menu is only accessible via a physically sealed internal
calibration switch positioned on the side of the register. Remove the drilled llister-head screw and use a
paper clip, small Allen wrench, or similar item to depress the switch.
VOLUME CALIBRATION
TEMP
MISC
Register functions carried out in W&M Mode are as follows
• Assign product name to each of a possible 10 products
• Add temperature compensation to each product, if required
• Enter meter K-Factor manually, if known
• Calculate meter K-Factor automatically using calibration prover volume
• Calibrate temperature probe in eld to known thermometer standard
• Input meter and E4000 serial number to be printed on reports
• Select units, gal/
o
F or liters/
o
C, and resolution of the display
• Select type of input pulse to match internal encoder or external device.
Note: Complete software details are listed in the E4000 Software Manual.
Note: See Section 5.4 for Calibrating E4000 Volumetric K-factor
Note: Always follow proper calibration procedures to avoid inuencing results.
• Recommend wet down test and discard results.
• Follow with 3 tests to determine repeatability.
• If repeatable, (within required specs) add results of the tests to get the average and put that
value in to adjust the k-factor. Then run verication test.
• If meter not repeatable, correct issues with installation, accessories, meter issues or vapor
issues before proceeding to item 2 above.