36
5.2 Conguring the E4000
1. At the root menu, remove the seal screw and rmly depress the Calibration switch. Press the [>] button
simultaneously using an allen key or stiff wire to enter the W&M mode indicated by ashing W&M. Press
[Enter] to continue.
ROUTE SUPERVISOR W&M
2. Select MISC from the calibration menu [Enter]
VOLUME CALIBRATION
TEMP
MISC
3. Use the push buttons [>] [^] to enter the Meter Serial Number [Enter]
METER SN
FA0516
4. Use the push buttons [>] [^] to enter the Register Serial Number [Enter]
REGISTER SN
E516EA
5. Select the correct volume units according to local standards. [Enter]
UNITS
VOLUME
GALLONS
6. Select the correct temperature units according to local standards. [Enter]
TEMPERATURE SCALE
Deg_C Deg_F