EasyManua.ls Logo

Red Seal Measurement Neptune RML2000 - Calibrating the E4000 Volumetric K-Factor; Pre-Calibration Steps and Shift Report

Red Seal Measurement Neptune RML2000
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
44
13. If the readings are acceptable, return to Prolink, Conguration, and under the Flow tab, change the
mass and vol ow cutoffs back to the original values (6.4000 lb/min Mass Flow Cutoff and 5.00000 gal/
min Vol Flow Cutoff). Note: This is very important. Failure to do this can result in the meter advancing the
E4000 totalizer, even though there is no ow through the meter.
14. Power down the system and disconnect the hardware from the meter transmitter to exit the Modbus
mode, and enter the pulsed output mode. Note: Until you exit the Modbus mode, no pulses will be sent
from the meter to the E4000 Electronic Register, and therefore no gallonage will be shown on the E4000
when pumping through the meter.
15. The zeroing operation is now complete, and when the system is powered back up, it will have a pulsed
output to the E4000. Open the valves, and calibrate the meter K factor using the E4000 Electronic Regis-
ter.
16. To ensure that you have successfully zeroed the meter, it is very important to test the meter at both a
high ow and low ow (82 GAL max. and 8 GAL min. for the RML2000). Note: While lling the volumet-
ric prover, do not allow the meter ow rate to decrease to less than 5 GPM, or the test results will not be
accurate. If there is no more then the allowed W&M spread (usually 1% maximum) between high and low
ow, the zero is acceptable. If greater than 1%, the meter should be zeroed again.
5.4 Calibrating the E4000 Volumetric K-Factor
Before calibration, see Section 5.3 for the RML2000 zeroing procedure.
NOTE: Software versions 08 and higher permit W&M inspectors to check calibration by viewing
both the gross and net volumes for the last delivery, without breaking the seal to enter the W&M
menus. If the ^ or > buttons are pressed after the delivery ticket is printed, but before the cancel
button is pressed to return to the SELECT PRODUCT screen, the screen will display both the gross
and net volumes. If calibration is required, it is recommended that a shift report be printed before
and after the calibration procedure, in order to reconcile the customers inventory. The shift
reports will show the totalizer readings at the start and completion of the calibration process. Any
quantity pumped through the meter during the calibration process will be added to the register
totalizer reading.
To print a shift report:
1. From the NEPTUNE root menu, press either > or ^ on the four button front panel, until ROUTE is
ashing, then press the START/ENTER button.
ROUTE SUPERVISOR W&M