45
2. Press > or ^ until REPORTS is ashing, then press the START/ENTER button.
DELIVERY VOLUME REPORTS
3. Press > or ^ until SHIFT is ashing, then press the START/ENTER button.
SHIFT CALIBRATION
4. Press the START/ENTER button when the display reads “PRINT END_OF_SHIFT REPORT?”
PRINT END_OF_SHIFT
REPORT?
5. The end of shift report will now print.
To Calibrate the Volumetric K-Factor:
1. Depress the Calibration switch and press the [>] button simultaneously to enter the W&M mode. When
W&M is ashing, press [Enter] to continue.
ROUTE SUPERVISOR W&M
2. Select VOL to enter the volume calibration menu, and press [Enter].
VOLUME CALIBRATION
TEMP
MISC