46
3. Select [^] the desired product code for the uid to be calibrated 0-9, and press [Enter].
SELECT PRODUCT CODE
0-9
4. Assign a name to the product code selected.
Scroll [>] [^] through the characters to enter the name, and then press [Enter].
PRODUCT NAME
PROPANE
5. Select [^] the required product class for temperature compensation
None, 505 LPG (US), 510 LPG (Canada), then press [Enter].
PRODUCT CLASS
505 LPG
6. Select the method of calibration to be used:
If the K-Factor is known and is being re-entered manually select MANUAL, press [Enter], and follow steps
7 and 8. If the K-Factor is not yet known select AUTO, then press [Enter] to skip to step 9.
Starting K-Factors* of 123 pulses/unit are programmed at the factory for Product 0.
CALIBRATION METHOD
MANUAL AUTO
* Note: Any product programmed with a K-Factor of 999999 cannot be selected for delivery in route
mode, and you cannot perform an AUTO calibration without rst doing a MANUAL calibration to
change the K-factor from 999999 to a starting K-factor of around 123.
7. Use [>] [^] to enter the known K-Factor, and press [Enter].
To calculate a new K-factor: NEW = OLD x REGISTER READING / CORRECTED PROVER READING
ENTER K-FACTOR
XXXXXX/UNIT