42
Transmitter Options Tab
8. While in Prolink, and Conguration, under the Flow tab, temporarily change the
mass and volume ow cutoffs to zero, and click Apply.
9. After the meter has been powered up for at least 30 minutes to warm up the
electronics, circulate product thru the meter for 5-10 minutes to remove vapor and
allow meter to reach a steady temperature. Close the valve downstream of the meter
rst, then the vapor return line valve, and nally close the valve upstream of the meter.
All three should be closed as quickly as possible to ensure the meter is completely lled
with LPG. See Figure 5.9. Disengage the truck PTO.
Note: This is a secure area. To access, go to your desktop, right click on
the Prolink icon, then click Properties. Under shortcut, add a space, then
“–Secure” at the end of the target le name (after .exe”).
Example: “…G700.exe” –Secure
FIGURE 5.9
FIGURE 5.8