Transmitter Options Tab
8. While in Prolink, and Conguration, under the Flow tab, temporarily change the
mass and volume ow cutoffs to zero, and click Apply.
9. After the meter has been powered up for at least 30 minutes to warm up the
electronics, circulate product thru the meter for 5-10 minutes to remove vapor and
allow meter to reach a steady temperature. Close the valve downstream of the meter
rst, then the vapor return line valve, and nally close the valve upstream of the meter.
All three should be closed as quickly as possible to ensure the meter is completely lled
with LPG. See Figure 5.9. Disengage the truck PTO.
Note: This is a secure area. To access, go to your desktop, right click on
the Prolink icon, then click Properties. Under shortcut, add a space, then
“–Secure” at the end of the target le name (after .exe”).
Example: “…G700.exe” –Secure