- 21 -
Error Possible Cause Corrective Action
The letter from a suction
channel is blinking and
the suction channel can-
not be opened.
In the single use mode only one
suction channel can be opened at
a time. There is currently a suction
channel already open.
&ORVHRQHRSHQVXFWLRQFKDQQHO¿UVWDQGWKHQRSHQ
the other.
Select the multi use mode, but to do so all suction
FKDQQHOVPXVW¿UVWEHFORVHG
It is not possible to
change between single
and multi use mode.
A suction channel is still open. Close all suction channels.
There is no audible
signal.
The audible signal is deactivated. Activate audible signal (Section. 5.6.2).
Suction hose cannot be
closed by pressing the
suction hose button.
An electrical appliance connected
to the QCB is still in use.
Switch the appliance connected to the QCB off, see
section 5.4.4.
The extraction unit re-
TXLUHV¿OWHUFOHDQLQJDW
frequent intervals.
The diameter of the suction tube is
too narrow, because a small suction
port was connected to the the suc-
tion point.
7KH¿QH¿OWHUXQLWLVVRGLUW\WKDWLW
is no longer possible to clean.
The system has a leak.
Use the original suction tubes and install a suction
port adapter if necessary (see accessories).
&KDQJH¿QH¿OWHUDGDSWHUVHFWLRQ
&KHFNWRPDNHVXUHWKDWWKHGXVWGUDZHUDQGWKH¿QH
¿OWHUXQLWKDYHEHHQUHSODFHGFRUUHFWO\DQGWKDWWKH
seals are not damaged.
Check to make sure that the dust area cover panel
KDVEHHQUHSODFHGFRUUHFWO\DQGLV¿[HGVHFXUHO\LQ
place.
Check to make sure the pinch valve seals are secure.
The suction does not
start when an electrical
appliance connected to
the QCB is activated.
The extraction unit is not switched
on.
Device fuse has been triggered.
QCB is not connected with the ex-
traction unit.
On / off threshold of the suction
channel is too high.
A connected remote control has
priority over the QCB for this suc-
tion channel.
Switch the extraction unit on (section 5.1).
Check fuses, if necessary replace fuse (section 6.3.1).
Connect the QCB to the extraction unit (section 4.6.1).
Adjust the on / off threshold (section 5.6.4).
Press the mouse scroll wheel from the relevant suc-
tion channel to give the QCB priority again.
The electrical appliance
connected to the QCB
cannot be operated.
QCB is not connected to a power
supply.
Connect the QCB to a power supply (section 4.6.1).
Suction begins, even
though no appliance
connected to the QCB
has been activated.
The on / off threshold from the suc-
tion channel is too low.
Check which suction channel is open, the relevant
letter will be shown in the display.
Adjust the on / off threshold from that particular suc-
tion channel (Section 5.6.4).
The suction does not
stop when the electrical
appliance is switched
off.
The electrical appliance was not
connected to the QCB.
The on / off threshold is set too low.
A connected remote control has
priority over the QCB for this suc-
tion channel.
Connect the electrical appliance to the QCB (section
4.6.2).
Adjust the on / off threshold (section 5.6.4).
Press the mouse scroll wheel from the relevant suc-
tion channel to give the QCB priority again.
The fault “Err14” ap-
pears when program-
ming the on / off thresh-
old for a handpiece.
The rpm you have selected on the
handpiece means the difference
between stand-by power and oper-
ating current is too low.
Program a higher rpm on the handpiece to activate
the on / off threshold.
,WLVQRWSRVVLEOHWR¿QG
a satisfactory setting for
the on / off threshold.
Some older handpieces do not
SURYLGHVXI¿FLHQWVLJQDOIRUWKHDX-
tomatic switch on.
Switch the suction channel on manually (see section
5.4.3).
EN