English 15
Problem/possible cause
Solution
Mask may be fitted incorrectly. Make sure your mask is fitted correctly. See your mask user
guide for fitting instructions or use the Mask Fit function to
check your mask fit and seal.
Displays messsage: Tubing blocked, check your tubing
Air tubing may be blocked. Check the air tubing and remove any blockages. Press the
dial to clear the message and then press Start/Stop to
restart the device.
Displays message: Leak detected, check your system setup and all connections
Mask may be fitted incorrectly. Make sure your mask is fitted correctly. See your mask user
guide for fitting instructions or use the Mask Fit function to
check your mask fit and seal.
Displays message: Low MV detected
Minute ventilation level has dropped below the pre-set
alarm level.
Contact your care provider.
Displays message: Apnea detected
The device detects an apnea that has exceeded the pre-set
alarm.
Breath normally to disable the alarm.
If the problem persists contact your care provider.
Displays message: Low Sp0
2
detected
SpO
2
has dropped below the pre-set alarm level. Check the attachment of the sensor.
If the problem persists contact your care provider.
Displays message: No SpO
2
data, check your oxi sensor attachment to module/finger
Oximeter sensor is not attached properly. Ensure that the oximeter sensor is attached properly to the
module and to your finger.
Oximeter sensor may be faulty. If the message appears repeatedly, the oximeter sensor
might be faulty. Replace the oximeter.
Displays message: Non-vented mask detected, use vented mask or unblock your mask vents
Non-vented mask is used. Only use a vented mask.
Mask vents might be blocked. Check if you have sufficient venting. Unblock mask vents if
necessary.
Expiratory pressure (EPAP) may be set too low. Talk to your care provider about your settings.
Displays message: System fault, refer to user guide, Error 004
Device may have been left in a hot environment. Allow to cool before re-use. Disconnect the power supply
and then reconnect it to restart the device.
Air filter may be blocked. Check the air filter and replace it if there are any blockages.
Disconnect the power supply and then reconnect it to restart
the device.

Table of Contents

Other manuals for ResMed AirCurve 10 ST-A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed AirCurve 10 ST-A and is the answer not in the manual?

ResMed AirCurve 10 ST-A Specifications

General IconGeneral
BrandResMed
ModelAirCurve 10 ST-A
CategoryMedical Equipment
LanguageEnglish

Related product manuals