English 19
Technical specifications
AC input range: 100–240V, 50–60Hz 1.0–1.5A, Class II
115V, 400Hz 1.5A, Class II (nominal for aircraft use)
Typical power consumption: 53W (57VA)
+41°F to +95°F (+5°C to +35°C)
Note: The air flow for breathing produced by this therapy
device can be higher than the temperature of the room.
Under extreme ambient temperature conditions
(104°F/40°C) the device remains safe.
10 to 95% relative humidity, non-condensing
Sea level to 8,500' (2,591 m); air pressure range 1013 hPa to
738 hPa
Storage and transport temperature:
-4°F to +140°F (-20°C to +60°C)
Storage and transport humidity:
5 to 95% relative humidity, non-condensing
Electromagnetic compatibility
The AirCurve 10 complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC
60601-1-2:2014, for residential, commercial and light industry environments. It is recommended that mobile communication
devices are kept at least 1 m away from the device.
For further details see the Guidance and manufacturer’s declaration – electromagnetic emissions and immunity section.
IEC 60601-1:2005 classification
Class II (double insulation), Type BF, Ingress protection IP22.
Internally located at device outlet, analog gauge pressure
type, -5 to +45 cm H
2
O (-5 to +45 hPa)
Internally located at device inlet, digital mass flow type, -70
to +180 L/min
Maximum single fault steady pressure
Device will shut down in the presence of a single fault if the steady state pressure exceeds:
30 cm H
2
O (30 hPa) for more than 6 sec or 40 cm H
2
O (40 hPa) for more than 1 sec.
Pressure level measured according to ISO 80601-2-70:2015 (CPAP mode):
SlimLine: 25 dBA with uncertainty of 2 dBA
25 dBA with uncertainty of 2 dBA
SlimLine or Standard and humidification: 27 dBA with uncertainty of 2 dBA
Power level measured according to ISO 80601-2-70:2015 (CPAP mode):
SlimLine: 33 dBA with uncertainty of 2 dBA
33 dBA with uncertainty of 2 dBA
SlimLine or Standard and humidification: 35 dBA with uncertainty of 2 dBA
Declared dual-number noise emission values in accordance with ISO 4871:1996.
Low (nominal 54 dBA), Medium (nominal 60 dBA), High (nominal 73 dBA)