English 23
Device without humidification and SlimLine air tubing / Device with humidification and SlimLine air tubing
2
16 0.5 / 0.5 0.5 / 0.5 0.8 / 0.8
Pressure accuracy - bilevel
Maximum dynamic pressure variation according to ISO 80601-2-70:2015.
Device without humidification and Standard air tubing / Device with humidification and Standard air tubing
Inspiratory pressure (cm H
2
O [hPa]) (Means, Standard Deviations)
-0.22, 0.01
-0.22, 0.01
-0.24, 0.01
-0.29, 0.03
-0.26, 0.02
-0.14, 0.02
-0.23, 0.01
-0.28, 0.01
-0.34, 0.01
-0.38, 0.02
-0.40, 0.03
-0.34, 0.03
Expiratory pressure (cm H
2
O [hPa]) (Means, Standard Deviations)
-0.14, 0.01 /
-0.27, 0.01
-0.16, 0.01 /
-0.29, 0.02
-0.11, 0.10 /
-0.34, 0.02
-0.16, 0.05 /
-0.33, 0.01
-0.17, 0.05 /
-0.33, 0.02
0.04, 0.17 /
-0.21, 0.01
-0.16, 0.01 /
-0.25, 0.01
-0.20, 0.01 /
-0.33, 0.02
-0.20, 0.05 /
-0.35, 0.01
-0.21, 0.05 /
-0.38, 0.02
-0.23, 0.08 /
-0.38, 0.02
-0.27, 0.01 /
-0.37, 0.01
-0.26, 0.02 /
-0.34, 0.01
-0.25, 0.01 /
-0.38, 0.01
-0.29, 0.01 /
-0.43, 0.02
-0.31, 0.01 /
-0.45, 0.03
-0.13, 0.23 /
-0.31, 0.01
Device without humidification and SlimLine air tubing / Device with humidification and SlimLine air tubing
Inspiratory pressure (cm H
2
O [hPa]) (Means, Standard Deviations)
-0.52, 0.01
-0.53, 0.02
-0.53, 0.01
-0.54, 0.02
-0.51, 0.02
-0.18, 0.02
-0.51, 0.01
-0.54, 0.01
-0.56, 0.01
-0.58, 0.02
-0.60, 0.03
-0.25, 0.02
-0.52, 0.01
-0.58, 0.01
-0.62, 0.01
-0.67, 0.02
-0.69, 0.02
-0.40, 0.02
Expiratory pressure (cm H
2
O [hPa]) (Means, Standard Deviations)
-0.28, 0.01 /
-0.43, 0.01
-0.30, 0.03 /
-0.50, 0.01
-0.30, 0.01 /
-0.54, 0.01
-0.33, 0.01 /
-0.58, 0.01
-0.34, 0.01 /
-0.60, 0.02
-0.27, 0.01 /
-0.30, 0.01
-0.24, 0.02 /
-0.37, 0.01
-0.29, 0.02 /
-0.47, 0.01
-0.35, 0.01 /
-0.55, 0.01
-0.38, 0.01 /
-0.62, 0.02
-0.42, 0.02 /
-0.66, 0.01
-0.33, 0.01 /
-0.36, 0.01
0.05, 0.21 /
-0.38, 0.01
-0.31, 0.02 /
-0.50, 0.02
-0.37, 0.02 /
-0.57, 0.02
-0.43, 0.02 /
-0.65, 0.02
-0.48, 0.02 /
-0.68, 0.02
-0.43, 0.02 /
-0.45, 0.01
Note: The table above is based on data that covers between 60.1 and 88.8% of the inspiratory phase and 66.1 and 93.4% of the expiratory
phase durations. These data time slots start immediately after the initial transient overshoot/undershoot periods and end at the point that
flow diminishes to an equivalent absolute value of its starting point, towards the end of the breath phases (this corresponds to the % ranges
of values given immediately above).