English 15
CAUTION
Use only ResMed parts and accessories with the device. Non-ResMed parts may reduce the
effectiveness of the treatment and/or damage the device.
Use only vented masks recommended by ResMed or by the prescribing doctor with this
device. Fitting the mask without the device blowing air can result in rebreathing of exhaled
air. Make sure that the mask vent holes are kept clear and unblocked to maintain the flow
of the fresh air into the mask.
Be careful not to place the device where it can be bumped or where someone is likely to
trip over the power cord.
Blocking the air tubing and/or air inlet of the device while in operation could lead to
overheating of the device.
Keep the area around the device dry, clean and clear of anything (eg, clothes or bedding)
that could block the air inlet or cover the power supply unit.
Do not place the device on its side as water might get into the device.
Incorrect system setup may result in incorrect mask pressure reading. Ensure the system is
correctly set up.
Do not use bleach, chlorine, alcohol, or aromatic-based solutions, moisturising or
antibacterial soaps or scented oils to clean the device, the humidifier or air tubing. These
solutions may cause damage or affect the humidifier performance and reduce the life of
the products.
If you use the humidifier, always place the device on a level surface lower than your head
to prevent the mask and air tubing from filling with water.
Leave the humidifier to cool for ten minutes before handling to allow the water to cool and
to make sure that the humidifier is not too hot to touch.
Make sure that the humidifier is empty before transporting the device.
Technical specifications
Units are expressed in cm H
2
O and hPa. 1 cm H
2
O is equal to 0.98 hPa.
90W power supply unit
AC input range: 100–240V, 50–60Hz 1.0–1.5A, Class II
115V, 400Hz 1.5A, Class II (nominal for aircraft use)
DC output: 24V
3.75A
Typical power consumption: 53W (57VA)
Peak power consumption: 104W (108VA)
Environmental conditions
Operating temperature: +5°C to +35°C
Note: The air flow for breathing produced by this therapy
device can be higher than the temperature of the room.
Under extreme ambient temperature conditions (40°C) the
device remains safe.
Operating humidity: 10 to 95% relative humidity, non-condensing
Operating altitude: Sea level to 2,591 m; air pressure range 1013 hPa to
738 hPa
Storage and transport temperature: -20°C to +60°C
Storage and transport humidity: 5 to 95% relative humidity, non-condensing

Other manuals for ResMed AirCurve 10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed AirCurve 10 and is the answer not in the manual?

ResMed AirCurve 10 Specifications

General IconGeneral
Sound Level26 dBA
Ramp FeatureYes
Device TypeBi-Level Positive Airway Pressure Device
Weight1248 g
HumidificationIntegrated humidifier
ConnectivitySD card
Mode of OperationS
Ramp Time0 to 45 minutes
Data StorageSD card
DisplayColor LCD screen
CompatibilityCompatible with masks
Power Supply100–240V AC, 50–60Hz, 2.0A