EasyManuals Logo

ResMed AirCurve 10 User Manual

ResMed AirCurve 10
22 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
16
Electromagnetic compatibility
The AirCurve 10 complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-
2:2007, for residential, commercial and light industry environments. It is recommended that mobile communication devices
are kept at least 1 m away from the device.
Information regarding the electromagnetic emissions and immunity of this ResMed device can be found on
www.resmed.com, on the Products page under Service and Support.
EN 60601-1:2006 classification
Class II (double insulation), Type BF, Ingress protection IP22.
Sensors
Pressure sensor: Internally located at device outlet, analogue gauge pressure
type, -5 to +45 cm H
2
O (-5 to +45 hPa)
Flow sensor: Internally located at device inlet, digital mass flow type, -70
to +180 L/min
Maximum single fault steady pressure
Device will shut down in the presence of a single fault if the steady state pressure exceeds:
30 cm H
2
O (30 hPa) for more than 6 sec or 40 cm H
2
O (40 hPa) for more than 1 sec.
Sound
Pressure level measured according to EN ISO 17510-1:2009 (CPAP mode):
SlimLine: 26.6 dBA with uncertainty of 2 dBA
Standard: 26.6 dBA with uncertainty of 2 dBA
Power level measured according to EN ISO 17510-1:2009 (CPAP mode):
SlimLine: 34.6 dBA with uncertainty of 2 dBA
Standard: 34.6 dBA with uncertainty of 2 dBA
Declared dual-number noise emission values in accordance with ISO 4871:1996.
Physical - device and humidifier
Dimensions (H x W x D): 116 mm x 255 mm x 150 mm
Air outlet (complies with ISO 5356-1:2004): 22 mm
Weight (device and cleanable humidifier): 1248 g
Housing construction: Flame retardant engineering thermoplastic
Water capacity: To maximum fill line 380 mL
Cleanable humidifier - material: Injection moulded plastic, stainless steel and silicone seal
Temperature
Maximum heater plate: 68°C
Cut-out: 74°C
Maximum gas temperature:
41°C
Air filter
Standard: Material: Polyester non woven fibre
Average arrestance: >75% for ~7 micron dust
Hypoallergenic: Material: Acrylic and polypropylene fibres in a polypropylene
carrier
Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron
dust
Aircraft use
ResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21,
category M) for all phases of air travel.

Other manuals for ResMed AirCurve 10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed AirCurve 10 and is the answer not in the manual?

ResMed AirCurve 10 Specifications

General IconGeneral
Sound Level26 dBA
Ramp FeatureYes
Device TypeBi-Level Positive Airway Pressure Device
Weight1248 g
HumidificationIntegrated humidifier
ConnectivitySD card
Mode of OperationS
Ramp Time0 to 45 minutes
Data StorageSD card
DisplayColor LCD screen
CompatibilityCompatible with masks
Power Supply100–240V AC, 50–60Hz, 2.0A