ReSound ONE 9 - Test Procedure for ReSound Coyote 6 CIC Instruments

ReSound ONE 9
14 pages
Print Icon
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Doc 0452940 rev. A
Page 14 of 14
Not subject to issue control when printed
1- Open Test Control. (DSA 6000)
2- Ensure Acoustical Calibration and Validation System
(Doc# 0202830) have been calibrated prior to testing
hearing instruments.
3- Wrap the canal tip with Fun-tak including the vent hole.
4- Leave vent on faceplate side open.
5- Check to see that the receiver tube is centered in the
coupler and the tip of the canal is flush with the coupler.
6- Connect programming cable and battery pill into device.
7- Place the device in the sound chamber box making sure
that it is on the blue area with the faceplate facing the
speaker.
Close the sound chamber and run the test.
Follow instructions on screen.
Test Procedure CIC