EasyManua.ls Logo

Respironics BiPAP Vision - System Final Test Setup

Respironics BiPAP Vision
244 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 7
Testing and Calibration
1062788 BiPAP Vision Service Manual © Respironics, Inc. 7-23
7.7.1 System Final Test Setup
CAUTION: To avoid damage to static-sensitive components, perform this procedure
an ESD-approved workstation. Ensure that trained service personnel
perform the system final test.
Figure 7-7 shows the initial equipment setup for the system final test. This
setup varies according the equipment used.
Figure 7-7: Initial Equipment Setup for System Final Test
1. Non-upgraded ventilators s/n 105999 and below: remove the
ventilator top enclosure, then connect the test cable from the service
laptop RS-232 port to the ventilator PCS connection (J3 on the PCS
board or the test cable labeled “PC”).
Newer (s/n 106000 and above) or upgraded ventilators: connect test
cable from service laptop to the RS-232 connector on the ventilator
back panel.
2. Connect patient tubing to the ventilator outlet. Connect a pressure
pick-off port to the tubing, then connect a flowmeter. Use another
length of patient tubing to connect the flowmeter outlet to a restrictor
valve. Adjust restrictor valve to create a small amount of circuit leak.
3. Use a tee fitting to connect the manometer, pressure pick-off port, and
ventilator pressure port together.
Flow analyzer
Flow restrictor
Certifier FA Plus
high flow module
Service laptop computer

Table of Contents

Related product manuals