8.2.2 Obsolete Repair Kits
lists repair kits that are no longer available and what repair kits should be used
instead.
RP-EPROM Upgrade Kit, non-
PAV
1000286 Software-only update with these features: O
2
% parameter
retained when switching modes, pressure increase from 10
to 15 cmH
2
O during Disconnect alarm condition, revised
alarm sounds (5-beep sequence for patient alarms, 3-beep
sequence for Check Vent alarms, soft click for out of range
setting attempt, original tone for Vent Inop alarms).
Includes:
• DC board Version 11 EPROM
• PC board Version 11 EPROM
• MC board Version 11 EPROM
RP-EPROM Upgrade Kit, PAV 1003524 Software-only update with these features: O
2
% parameter
retained when switching modes, pressure increase from 10
to 15 cmH
2
O during Disconnect alarm condition, revised
alarm sounds (5-beep sequence for patient alarms, 3-beep
sequence for Check Vent alarms, soft click for out of range
setting attempt, original tone for Vent Inop alarms).
Includes:
• DC board Version 11 EPROM
• PC board Version 11 EPROM
• MC board Version 11 EPROM
• Alarm B option PAL (not used for ventilators that already
have PAV)
Table 8-1: Upgrades for s/n 105999 and Below
Upgrade Part number Description
Table 8-2: Obsolete Repair Kits
Obsolete kit Substitution repair kit
Circulation fan muffler
French (P/N 1000741)
Spanish (P/N 1000738)
German (P/N 1000729)
Circulation fan muffler, international (P/N 1005618)
Circulation fan muffler, English, 3.5 A
rating (P/N 582155)
Circulation fan muffler, English, 4.0 A rating
(P/N 1041193)
Install when replacing 3.5-A fuses with 4.0-A fuses.
EPROM upgrade kit (P/N 582180) EPROM upgrade kit, non-PAV, s/n 105999 and below