WARNING
ALWAYS unplug the unit from the electrical
outlet when cleaning the unit.
It is recommended to clean the Heating
Unit a minimum of 2 times per year, or
more frequently as needed — especially if
not using with HydraHeat packs.
CAUTION
DO NOT USE BLEACH or any cleaner with
HIGH CHLORINE content.
NOTE: Chlorine in regular tap water may
be present in high enough concentrations
(4ppm or more) to damage your unit.
If you suspect high levels of chlorine in
your HydraTherm Heating Unit, Richmar
recommends the use of distilled water or
the addition of a dechlorinator.
NOTE: DO NOT USE additives such as
herbal teas, essential oils, etc., may damage
the components of the Heating Unit. It is
the manufacturers recommendation to use
the HydraTherm Heating Unit with clean
water and using the manufacturer supplied
HydraHeat packs.
The interior of the unit should be cleaned
using a non-chlorinated and non-abrasive
cleaner. Check to make sure your cleaner
has a low amount or no chlorine content
in your cleaner and make sure that any
residue is throughly rinsed away with water.
To aid in the removal of deposits on
the heating element, pour a solution of
vinegar and water in a 1:1 ratio into the
heating element well, using no more than
required to submerge the heating element.
Throughly rinse the water tank with clean
water to remove dissolved deposits.
NOTE: Failure to maintain your equipment
may result in voiding the warranty.
CLEANING STEPS
STEP 1
Turn device o to let the
water and packs cool to
room temperature.
STEP 2
Remove packs from device and use
sponge to clean with cleaning solution.
STEP 3
Remove Divider (or Rack) System to
access bottom well of the
HydraTherm™ tank.
STEP 4
Remove any debris (e.g. fabric lint) from the
bottom of the tank, including the well around
the heating element, before beginning to
drain the device.
CAUTION
DO NOT attempt to clean the debris from
a tank with hot water. Please allow to cool
to room temperature.
STEP 5
Deluxe HydraTherm™
1. Position hose over wastewater receptacle.
2. Open valve by turning handle parallel
with hose.
3. Change the device mode from
HEAT to DRAIN.
4. Toggle the power switch to ON.
Water will begin emptying from the hose.
5. Once tank is empty, check the filter for
debris by unscrewing to remove, clean
out the debris and replace before closing
the valve.
Standard HydraTherm™
1. Position hose over wastewater
receptacle.
2. Open valve by turning handle
parallel with hose. Water will
begin to drain from the hose.
STEP 6
1. When tank is empty, close valve
by turning handle perpendicular
to the drain hose.
2. Soak sponge in cleaning solution
and wipe all surfaces of the
HydraTherm™ tank and dividers.
3. Rinse sponge and soak in water,
and then rinse tank.
4. Remove any water and cleaning solution
using appropriate draining method above.
STEP 7
Refer to Quick Start Guide or Operation
Manual for instructions on how to properly
fill your device.
CLEANING INSTRUCTIONS
10
Device switch OFF
Valve CLOSED
Valve OPEN
Device mode: DRAIN
DRAIN HEAT
Power switch: ON
Valve OPEN
Valve CLOSED