Chapter 5 Specifications  RIGOL 
DG1000Z User’s Guide  5-7 
 
Temperature Range 
Operating: 0℃ to 50℃ 
Non-Operating: -40℃ to 70℃ 
Cooling Method  Cooling by fans compulsively  
Humidity Range 
Less than 30℃: ≤95% Relative Humidity (RH) 
30℃ to 40℃: ≤75% Relative Humidity (RH) 
40℃ to 50℃: ≤45% Relative Humidity (RH) 
Altitude 
Operating: Less than 3000 meters 
Non-Operating: Less than 15,000 meters 
(W×H×D) 
261.5mm×112mm×318.4mm 
Weight 
with package: 4.5kg 
Interfaces  USB Host, USB Device, LAN 
Interval 
Recommend calibration interval is one year  
Authentication Information 
EMC 
In line with 
EN61326-1:2006 
 
IEC 61000-3-2:2000 
±4.0kV (Contact Discharge) 
±4.0kV (Air Discharge) 
IEC 61000-4-3:2002 
3V/m (1.4GHz to 2GHz) 
1V/m (2.0GHz to 2.7GHz) 
IEC 61000-4-4:2004  1kV power lines 
IEC 61000-4-5:2001 
0.5kV (Phase to Neutral) 
0.5kV (Phase to PE) 
1kV (Neutral to PE) 
EC 61000-4-11:2004 
Voltage dip: 
0%UT during half cycle 
0%UT during 1 cycle 
70%UT during 25 cycle 
Short interruption: 
0%UT during 1 cycle 
Electrical Safety 
USA: UL 61010-1:2012, 
Canada: CAN/CSA-C22.2 No. 61010- 1-2012 
EN 61010-1:2010