EasyManua.ls Logo

Rigol DHO1000 Series

Rigol DHO1000 Series
39 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Channel
Vertical
Scale
Test Result
Limit Pass/Fail
Vavg1 Vavg2
Calculatio
n Result
[1]
2 mV/div ≤2%
5 mV/div ≤2%
10 mV/div ≤2%
20 mV/div ≤2%
50 mV/div ≤2%
100 mV/div ≤2%
200 mV/div ≤2%
500 mV/div ≤2%
1 V/div ≤2%
2 V/div ≤2%
5 V/div ≤2%
10 V/div ≤2%
CH4
1 mV/div ≤2%
2 mV/div ≤2%
5 mV/div ≤2%
10 mV/div ≤2%
20 mV/div ≤2%
50 mV/div ≤2%
100 mV/div ≤2%
200 mV/div ≤2%
500 mV/div ≤2%
1 V/div ≤2%
2 V/div ≤2%
5 V/div ≤2%
10 V/div ≤2%
NOTE
[1]: The calculation formula is |(Vavg1 - Vavg2) - (Vout1 - Vout2)|/Full Scale x 100%; wherein,
Vout1 is 3 x the current vertical scale and Vout2 is -3 x the current vertical scale.
DC Offset Accuracy Test Record Form
Channel
Setting
Vertical
Scale
Offset Test Result Min. Max.
CH1, 20
MHz BW
500 μV/div 1 V -1.01 V -982.95 mV
500 μV/div 0 V -2.05 mV 2.05 mV
500 μV/div -1 V -982.95 mV 1.01 V
200 mV/div 10 V -10.17 V -9.92 V
200 mV/div 0 V -22.00 mV 22.00 mV
200 mV/div -10 V 9.92 V 10.17 V
1 V/div 20 V -20.30 V -19.69 V
Appendix: Test Record Form
DHO1000 Performance Verification
30
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.