RIGOL         Chapter 2 Front Panel Operations 
2-8  DL3000 User’s Guide 
 
Figure 2-6 Main Interface of CV Mode 
 
In CV mode, you need to set the following parameters: Voltage, Range, V_Limit, 
C_Limit. 
 
Set Voltage 
The constant voltage value in CV mode. The default unit for the voltage is V. 
(1)  In the main interface, press Voltage to switch the parameter focus to 
"Voltage". 
(2)  Enter a value by using the numeric keys, arrow keys, or the knob. For the 
setting methods, refer to the descriptions of the built-in help information or 
refer to the descriptions in "Parameter Setting Method". 
 
 
Set Range 
Working range for the voltage in CV mode. Two ranges are available: low range 
(0-15 V) and high range (0-150 V). 
(1)  In the main interface, press Range to switch the parameter focus to 
"Range". 
(2)  Press Range to switch the voltage working range. You can also use the 
Left/Right arrow key or the knob to switch it.  
 
Note: 
  The low range provides better resolution and accuracy at low voltage 
settings. 
  If the set voltage value is greater than the maximum value of the low range, 
you must select a high range. 
 
 
Before switching the voltage range, please disable the channel input 
to avoid causing damage to the instrument or the DUT. 
 
 
In CV mode, when the set load voltage is greater than the voltage output 
from the DUT, 
 occurs to the DUT.