Chapter 2 Front Panel Operations RIGOL
DL3000 User’s Guide 2-31
CC Toggled Mode (Tog)
In Tog mode, when you enable the transient test, the load current will toggle for one
time between Level A and Level B once the load receives a trigger signal, as shown in
Figure 2-22.
Figure 2-22 CC Toggled Mode (Tog)
Operation Procedures:
1. Connect the DUT and the channel input terminals of the load
Turn off the instrument, as shown in Figure 2-2, connect the DUT and the
channel input terminals on the front panel of the load.
While making a connection, the positive polarity of the load should
be connected to the (+) terminal of the channel output, and the
negative polarity of the load to the (-) terminal of the channel
output. A misconnection with the terminals may cause damage to
the instrument or the DUT.
2. Power on the instrument
After the instrument is connected to the power source, press the Power key
on the front panel to power on the instrument.
3. Set channel parameters in the guide interface
Press Tog to enter the Tog guide interface, as shown in Figure 2-23
. In the
guide interface, you can configure parameters in the parameter configuration
list and view the configuration diagram at the left of the configuration list in the
real-time manner.
TRIG