RIGOL         Chapter 2 Front Panel Operations 
2-54  DL3000 User’s Guide 
Set P_Start 
The power value when the load starts in OPP test mode. The default unit for the 
value is W. 
(1)  In the guide interface, press the Up/Down arrow key to switch the 
parameter focus to "P_Start". 
(2)  Enter a value by using the numeric keys. For setting methods, refer to the 
descriptions in "Parameter Setting Method". 
 
Set P_Step 
The step power in OPP test mode. The default unit for the value is W. 
(1)  In the guide interface, press the Up/Down arrow key to switch the 
parameter focus to "P_Step". 
(2)  Enter a value by using the numeric keys. For setting methods, refer to the 
descriptions in "Parameter Setting Method". 
 
Set Delay_Step 
The time interval during which the power moves from step to step in OPP test 
mode. The default unit for the value is ms. 
(1)  In the guide interface, press the Up/Down arrow key to switch the 
parameter focus to "Delay_Step".  
(2)  Enter a value by using the numeric keys. For setting methods, refer to the 
descriptions in "Parameter Setting Method". 
 
Set OPP_V 
The protection voltage in the OPP test mode. The default unit for the value is V. 
(1)  In the guide interface, press the Up/Down arrow key to switch the 
parameter focus to "OPP_V". 
(2)  Enter a value by using the numeric keys. For setting methods, refer to the 
descriptions in "Parameter Setting Method". 
 
Set P_Max 
The maximum value of the protection power in the OPP test mode. The default 
unit for the value is W. 
(1)  In the guide interface, press the Up/Down arrow key to switch the 
parameter focus to "P_Max". 
(2)  Enter a value by using the numeric keys. For setting methods, refer to the 
descriptions in "Parameter Setting Method". 
 
Set P_Min 
The minimum value of the protection power in the OPP test mode. The default 
unit for the value is W. 
(1)  In the guide interface, press the Up/Down arrow key to switch the 
parameter focus to "P_Min". 
(2)  Enter a value by using the numeric keys. For setting methods, refer to the 
descriptions in "Parameter Setting Method".