RIGOL Chapter 2 Front Panel Operations
2-54 DL3000 User’s Guide
Set P_Start
The power value when the load starts in OPP test mode. The default unit for the
value is W.
(1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to "P_Start".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set P_Step
The step power in OPP test mode. The default unit for the value is W.
(1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to "P_Step".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set Delay_Step
The time interval during which the power moves from step to step in OPP test
mode. The default unit for the value is ms.
(1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to "Delay_Step".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set OPP_V
The protection voltage in the OPP test mode. The default unit for the value is V.
(1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to "OPP_V".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set P_Max
The maximum value of the protection power in the OPP test mode. The default
unit for the value is W.
(1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to "P_Max".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".
Set P_Min
The minimum value of the protection power in the OPP test mode. The default
unit for the value is W.
(1) In the guide interface, press the Up/Down arrow key to switch the
parameter focus to "P_Min".
(2) Enter a value by using the numeric keys. For setting methods, refer to the
descriptions in "Parameter Setting Method".

Table of Contents

Other manuals for Rigol DL3000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DL3000 and is the answer not in the manual?

Rigol DL3000 Specifications

General IconGeneral
BrandRigol
ModelDL3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals