EasyManua.ls Logo

Rigol DM3058 - Page 131

Rigol DM3058
264 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL
Programming Guide for DM3058
5-15
Parameters:
Name
Type
Range of Values
Default
<range>
Consecutive
Integer
20 Hz~1 MHz
switching to
autorange
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium
Explanations:
<range> may be 20 Hz~1 MHz (unit is Hz). MAX = 1 000 000, MIN = 20, DEF is
autorange.
<resolution> (unit is Hz); in default setting, the rate is medium.
8. CONFigure:FRESistance
Command Format:
CONFigure:FRESistance
[{<range>|AUTO|MIN|MAX|DEF} [,{<resolution>|MIN|MAX|DEF}]]
Function:
The command resets all the 4-wire resistance measuring parameters and trigger
parameters as defaults, and sets 4-wire resistance measuring function.
Parameters:
Name
Type
Range of Values
Default
<range>
Discrete
{200 Ω | 2 kΩ | 20 kΩ | 200
| 2 MΩ | 10 MΩ | 100 MΩ}
2
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium
Explanations:
<range> may be 200 Ω (MIN), 2 , 20 , 200 , 2 , 10 , 100 (MAX),
AUTO (autorange) (unit is Ω). MAX = 100 000 000, MIN = 200.
<resolution> (unit is Ω); in default setting, the rate is medium.
9. CONFigure:PERiod
Command Format:
CONFigure:PERiod[{<range>|MIN|MAX|DEF}[,{<resolution>|MIN|MAX|DEF
}]]