RIGOL
Programming Guide for DM3058
5-28
1. MEASure:CAPacitance?
Command Format:
MEASure:CAPacitance?
[{<range>|AUTO|MIN|MAX|DEF} [,{<resolution>|MIN|MAX|DEF}]]
Function:
The command resets all the capacitance measuring parameters and trigger
parameters as defaults, and sets capacitance measuring function, gets the
measuring results and sends them to output buffer of instrument.
Parameters:
Name
Type
Range of Values
Default
<range>
Discrete
{2 nF | 20 nF | 200 nF| 2 µF
| 20 µF | 200 µF| 10000 µF}
200 nF
<resolution>
Discrete
{MIN|MAX|DEF}
Rate: medium
Explanations:
<range> may be 2 nF (MIN), 20 nF, 200 nF, 2 μF, 20 μF, 200 μF, 10000 μF (MAX),
AUTO (autorange) (unit is F). MAX = 0.01, MIN = 0.000 000 002.
<resolution> (unit is F); in default setting, the rate is medium.
Return Format:
The query returns the data in output buffer.
2. MEASure:CONTinuity?
Command Format:
MEASure:CONTinuity?
Function:
The command resets all the continuity measuring parameters and trigger
parameters as defaults, and sets continuity measuring function, get the
measuring results and send them to output buffer of instrument.
Explanations:
Enable continuity measuring function, and set the range as 2 kΩ.