1. SAMPle:SOURce
Command Format:
SAMPle:SOURce□{IMMediate|TIMer}
SAMPle:SOURce?
Function:
The command sets or queries the trigger type for sampling.
Explanations:
The command ―SAMPle:SOURce□{IMMediate|TIMer}‖ should be used together
with the command ―TRIGger:DELay‖ and ―SAMPle:TIMer‖. When sample
numbers are greater than 1, sample time can be determined; when sample
number is 1, the command is invalid.
SAMPle:SOURce?: Query the trigger type, including auto trigger (IMMediate) or
interval trigger (TIMer).
It is suggested that accurate range should be set by manual not by auto in all
trigger command.
Return Foramt:
The query returns the parameter related to trigger type: IMMediate or TIMer.
2. SAMPle:TIMer
Command Format:
SAMPle:TIMer□{<interval>|MIN|MAX}
SAMPle:TIMer?□[{MIN|MAX}]
Function:
The command sets or queries the sample interval time.
Parameters: