EasyManua.ls Logo

Rigol DM3058 - Page 164

Rigol DM3058
264 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL
Programming Guide for DM3058
5-48
1. [SENSe:]FUNCtion[:ON]
Command Format:
[SENSe:]FUNCtion[:ON]"<function>"
[SENSe:]FUNCtion[:ON]?
Function:
The command selects or queries measuring function.
Parameters:
Name
Type
Range of Values
Default
<function>
Discrete
{ 'volt:dc' | 'volt:ac' | 'curr:ac'
| 'curr:dc' | 'freq' | 'per' | 'cap'
| 'res' | 'fres' | 'cont' | 'diod' }
Required
parameter
Explanations:
[SENSe:]FUNCtion[:ON] "<function>": Select the measuring function for
instrument, the parameter enclosed by single quotation marks (‗‘) is a character
string related to measuring function. And the default is inexsitent, <function>
must be set. See table 5-2.
[SENSe:]FUNCtion[:ON]?: Query current measuring function.
Table 5-2 The parameters and their meanings
Parameter
Meaning
Parameter
Meaning
'volt:dc'
DCV
'cap'
capacitance
'volt:ac'
ACV
'res'
2WR
'curr:ac'
ACI
'fres'
4WR
'curr:dc'
DCI
'cont'
Continuity
'freq'
Frequency
'diod'
Diode
'per'
Period
Return Format:
The query returns the parameter related to current measuring function.
2. [SENSe:]CURRent:AC:NULL[:STATe]
Command Format: