1. TRIGger:COUNt
Command Format:
TRIGger:COUNt□{<count>|MIN|MAX|DEF|INFinity}
TRIGger:COUNt?□[{MIN|MAX}]
Function:
The command sets or queries the trigger times used in multimeter.
Parameters:
Explanations:
TRIGger:COUNt□{<count>|MIN|MAX|DEF}: Set the trigger times.
TRIGger:COUNt?□[{MIN|MAX}]: Query the trigger times.
The command is only available for remote operation, not for front panel
operation.
The command ―CONFigure‖ or ―MEASure?‖ can be used to set trigger times as 1.
It is suggested that the ―SENSe:<function>:RANGe:AUTO□OFF‖ should be used
to disable autorange function, the command ―SENSe:<function>:RANGe‖ ,
―CONFigure‖ or ―MEASure?‖ be used to set range by manual.
Return Format:
The query returns the trigger times used in multimeter.
2. TRIGger:DELay
Command Format:
TRIGger:DELay□{<seconds>|MIN|MAX|DEF}
TRIGger:DELay?□[{MIN|MAX}]
Function:
The command sets or queries the delay time between trigger signal and starting
measurement.