RIGOL
Programming Guide for DM3058
5-102
The command ―CONFigure‖ or ―MEASure?‖ can be used to set delay time.
It is suggested that the command ―SENSe:<function>:RANGe:AUTOOFF‖
should be used to disable autorange function, the command ―SENSe:<function>:
RANGe‖, ―CONFigure‖ or ―MEASure?‖ be uesd to set range by manual.
Return Format:
The query returns the enable state of trigger autodelay function.
4. TRIGger:SLOPe
Command Format:
TRIGger:SLOPe{POSitive|NEGative}
TRIGger:SLOPe?
Function:
The command sets or queries external trigger type, including rising edge
(POSitive) and falling edge (NEGative).
Explanations:
TRIGger:SLOPe{POSitive|NEGative}: Set the external trigger type as POSitive
or NEGative.
TRIGger:SLOPe?: Query the external trigger type.
It is suggested that the command ―SENSe:<function>:RANGe:AUTOOFF‖
should be used to disable autorange function, the command ―SENSe:<function>:
RANGe‖, ―CONFigure‖ or ―MEASure?‖ be uesd to set range by manual.
Return Format:
The query returns the external trigger type.
5. TRIGger:SOURce
Command Format:
TRIGger:SOURce{IMMediate|EXTernal|BUS}
TRIGger:SOURce?