Explanation:
This command is invalide for Dual Display.
Return Format:
The query returns the current period in the form of 9.185433e-05, and the unit is
s.
27. :measure:period
Command Format:
:measure:period□{<range>|MIN|MAX|DEF}
Function:
The command sets the measuring range of input voltage in period measurement.
Parameters:
Explanations:
The period range is 1 us~50 ms.
The Table 3-2 indicates the parameters and their ranges.
Example:
To set the measuring range of input voltage as 20 V:
:measure:period□DEF or :measure:period□2
28. :measure:period:range?
Command Format:
:measure:period:range?
Function:
The command queries the measuring range of AC voltage in period
measurement.