EasyManua.ls Logo

Rigol DM3058

Rigol DM3058
264 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL
Programming Guide for DM3058
3-60
Function:
The command queries and sets upper value in P/F calculation.
Parameters:
Name
Type
Range of Values
Default
<value>
consecutive real number
See table 3-11
0
Explanations:
:calculate:limit:upper?: Queriy the upper value of P/F operation.
:calculate:limit:upper<value>: Set
the upper value of P/F calculation. The
default range of <range> please refer to Table 3-11.
Return Format:
The query returns current upper value.
18. :calculate:limit:state
Command Format:
:calculate: limit:state?
:calculate: limit:state{ON|OFF}
Function:
The command queries the enable state of P/F operation.
Explanations:
:calculate: limit:state?: Enable or disable P/F operation.
:calculate: limit:state{ON|OFF}: Set the state of P/F operation.
Return Format:
The query returns the enable state of P/F operation: ON or OFF that denotes
turning P/F operation on or off.
.
Example:
Disable P/F operation:
:calculate:limit:stateOFF

Other manuals for Rigol DM3058