17
Specifications
All the specifications
[1]
can only be guaranteed when the instrument is operated
continuously for more than 30 minutes under the specified operation temperature.
Number of Channels
Model Number of Channels
DP932A 3
DP932U 3
DP932E 3
DC Output (0℃~40℃)
DC output (0℃~40℃)
Voltage/Current OVP/OCP
DP932A/
DP932U
CH1 0 to 32 V/0 to 3 A
1 mV to 35.2 V/1 mA to 3.3
A
CH2 0 to 32 V/0 to 3 A
1 mV to 35.2 V/1 mA to 3.3
A
CH3 0 to 6 V/0 to 3 A 1 mV to 6.6 V/1 mA to 3.3 A
DP932E
CH1 0 to 30 V/0 to 3 A 1 mV to 33 V/1 mA to 3.3 A
CH2 0 to 30 V/0 to 3 A 1 mV to 33 V/1 mA to 3.3 A
CH3 0 to 6 V/0 to 3 A 1 mV to 6.6 V/1 mA to 3.3 A
Internal Series/Parallel Mode
Internal series/parallel mode
Series mode voltage 64 V
Parallel mode current 6 A
Load Regulation Rate
Load regulation rate, ± (% of output + offset)
Voltage
[2]
<0.01%+2 mV
Specifications