EasyManua.ls Logo

Rigol DS1000CA Series - Page 71

Rigol DS1000CA Series
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL
Copyright ©1998-2008 RIGOL Technologies, Inc.
User’s Guide for DS1000CA Series
2-45
Trigger Holdoff
Use trigger Holdoff to stabilize a complex waveform, such as a pulse range. Holdoff
time is the oscilloscope’s waiting period before starting a new trigger. During Holdoff,
oscilloscope will not trigger until Holdoff ends. For instance: To trigger on the first
pulse on a group of them, users can set the holdoff time to Pulse cluster width.
Figure 2-55 Trigger Holdoff
To use trigger Holdoff:
1. Press the trigger MENU button to display the Menu.
2. Press Set Up key to display trigger set up menu.
3. Turn the multi function knob (
) to change Holdoff time until waveform is stable.
4. Pushing Trigger Hold off reset to reset the Holdoff time to its default value.
Time of holdoff
Trigger site
Trigger holdoff

Other manuals for Rigol DS1000CA Series

Related product manuals