EasyManua.ls Logo

Rigol DS1000CA Series - To Set up the Sampling System

Rigol DS1000CA Series
154 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL
Copyright ©1998-2008 RIGOL Technologies, Inc.
User’s Guide for DS1000CA Series
2-49
To Set Up the Sampling System
Figure 2-56 shows, the Acquire button at the MENU of the front panel.
Figure 2-56
The menu
Press the Acquire button, the interface menu shows as follows:
Figure 2-57
Table 2-34 The Acquire menu
Menu Settings Comments
Acquisition
Normal
Average
Peak Detect
Normal Acquisition mode.
Average Acquisition mode.
Peak Detect Acquisition mode.
Averages 2 to 256
Step by multiple of two. Set
average times from 2 to 256.
Sampling
Real-Time
Equ-Time
Real-time sampling mode.
Equivalent sampling mode.
SinX/X
ON
OFF
Set the insert mode is SinX or
Line.
Sa Rate Display sampling rate.
Acquire setup button

Other manuals for Rigol DS1000CA Series

Related product manuals