RIGOL
Copyright ©1998-2008 RIGOL Technologies, Inc.
User’s Guide for DS1000CA Series
5-1
Chapter 5 : Specifications
All specifications apply to the DS1000CA Series Oscilloscopes and a probe with the
Attenuation switch set to 10X unless noted otherwise. To meet these specifications,
two conditions must first be met:
1 The instrument must have been operating continuously for thirty minutes within
the specified operating temperature.
2 Must perform the Self Cal operation, accessible through the Utility menu, if the
operating temperature changes by more than 5.
All specifications are guaranteed unless noted “typical”.
Specifications
Acquisition
Sampling Modes Real-Time Equivalent
Sampling Rate 2GSa/s
[1]
50GSa/s
[2]
Averages N time acquisitions, all channels simultaneously, N is
selectable from 2, 4, 8, 16, 32, 64, 128 and 256
Inputs
Input Coupling DC, AC, GND
Input Impedance 1M±2%, in parallel with 15pF±3pF
50±2%
[3]
Probe Attenuation Factors 1X, 5X, 10X, 50X, 100X, 500X, 1000X
Maximum Input Voltage
300V (DCAC Peak, 1M input impedance)
5V (DC+AC Peak, 50 input impedance)
[3]
Time delay between channel
(typical)
500ps

Other manuals for Rigol DS1000CA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1000CA Series and is the answer not in the manual?

Rigol DS1000CA Series Specifications

General IconGeneral
BrandRigol
ModelDS1000CA Series
CategoryTest Equipment
LanguageEnglish

Related product manuals