Page #44 background imageLoading...
Page #44 background image
RIGOL Command Systems
Programming Guide for DS1000B Series
Example:
:TRIG:HOLD 0.0001 Set the holdoff time as 100μs.
:TRIG:HOLD? Return 1.000e-004.
9. :TRIGger:STATus?
Query Format:
:TRIGger:STATus?
Function:
The command queries the current status of the oscilloscope. The status may be
RUN, STOP, T’ D, WAIT, SCAN or AUTO.
Returned Format:
The query returns RUN or STOP, T’D, WAIT, AUTO.
10. :Trig%50
Command Format:
:Trig%50
Function:
The command sets the trigger level at the vertical midpoint of the signal amplitude.
11. :FORCetrig
Command Format:
:FORCetrig
Function:
The command will produce a trigger signal to force the oscilloscope trigger and
to display a waveform when there is no suitable trigger condition.
NOTE: It is mainly applicable to the “Normal” and “Single” trigger modes.
12. :SINGLE

Other manuals for Rigol DS1204B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1204B and is the answer not in the manual?

Rigol DS1204B Specifications

General IconGeneral
BrandRigol
ModelDS1204B
CategoryTest Equipment
LanguageEnglish

Related product manuals