EasyManuals Logo
Home>Rigol>Test Equipment>DS7024

Rigol DS7024 User Programming Manual

Rigol DS7024
259 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #209 background imageLoading...
Page #209 background image
Chapter 2 Command System RIGOL
MSO7000/DS7000 Programming Guide 2-195
:TRIGger:SHOLd:CLEVel
Syntax
:TRIGger:SHOLd:CLEVel <level>
:TRIGger:SHOLd:CLEVel?
Description
Sets or queries the trigger level of the clock source. The unit is the same as that of the
current amplitude.
Parameter
Name
Type
Range
Default
<level>
Real
(-5 × VerticalScale - OFFSet) to
(5 × VerticalScale - OFFSet)
0 V
Remarks
For VerticalScale, refer to the :CHANnel<n>:SCALe command; for OFFSet, refer to
the :CHANnel<n>:OFFSet command.
Return
Format
The query returns the trigger level of the clock source in scientific notation.
Example
:TRIGger:SHOLd:CLEVel 0.05 /*Sets the trigger level of the clock source to 50 mV.*/
:TRIGger:SHOLd:CLEVel? /*The query returns 5.000000E-2.*/

Table of Contents

Other manuals for Rigol DS7024

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS7024 and is the answer not in the manual?

Rigol DS7024 Specifications

General IconGeneral
BrandRigol
ModelDS7024
CategoryTest Equipment
LanguageEnglish

Related product manuals