Chapter 2 Command System RIGOL
MSO1000Z/DS1000Z Programming Guide 2-3
:SINGle
Syntax
:SINGle
Description
Set the oscilloscope to the single trigger mode. This command is equivalent to any of the
following two operations: pressing the SINGLE key at the front panel and sending
the :TRIGger:SWEep SINGle command.
Explanation
 In the single trigger mode, the oscilloscope triggers once when the trigger conditions
are met and then stops.
 When the waveform record function is enabled or during the playback of the
recorded waveform, this command is invalid.
Related
Commands
:TFORce
:RUN
:STOP
:TFORce
Syntax
:TFORce
Description
Generate a trigger signal forcefully. This command is only applicable to the normal and
single trigger modes (see the :TRIGger:SWEep command) and is equivalent to pressing
the FORCE key in the trigger control area at the front panel.