142 • PID Relative Response Feature EAGLE 2 Operators Manual
2. With the PID Sensor Relative Response Screen displayed, press and
release AIR YES. A list of gases will appear on the screen with EXIT
at the top of the list.
There are multiple screens of gases. The following is the complete list of
factory defined gases along with their detection ranges, low alarm, high
alarm, STEL, and TWA settings. Table 14 is a list of the low range values
and Table 15 is a list of the high range values.
Table 14: Low Range Relative Response Gas List
Target Gas Detection
Range (ppm)
Alarm 1 Factory
Setting (ppm)
Alarm 2 Factory
Setting (ppm)
STEL
(ppm)
TWA
(ppm)
Acetone 0-30.00 5.00 7.50 OFF OFF
Benzene 0-25.00 0.50 2.50 2.50 0.50
Diesel Fuel
NO 1
0-40.00 2.00 3.00 OFF OFF
Ethanol 0-400.00 10.0 15.0 OFF OFF
Gasoline 0-50.00 3.00 5.00 OFF OFF
Isobutylene 0-50.00 4.30 6.00 OFF OFF
Isopropanol 0-200.00 2.0 4.0 OFF OFF
JP-5 Fuel 0-30.00 1.40 2.10 OFF OFF
Methyl Ethyl
Ketone
0-40.00 2.00 3.00 OFF OFF
Toluene 0-25.00 0.50 1.50 OFF OFF
N-Hexane 0-200.00 5.0 10.0 OFF OFF
Propylene 0-50.00 5.00 7.50 OFF OFF
Styrene 0-20.00 0.20 0.40 OFF OFF
Tetrachloro-
ethylene
0-30.00 0.24 1.00 OFF OFF
Trichloro-
ethylene
0-30.00 0.50 1.00 OFF OFF
Vinyl Chloride 0-100.00 1.0 5.0 5.0 1.0
>EXIT
ACETONE
BENZENE
DIESEL FUEL NO 1
ETHANOL
GASOLINE

Table of Contents