296 Alarms EAGLE 2 Operator’s Manual
8. Move the cursor next to the oxygen channel. Press and release POWER
ENTER RESET.
9. A mode select screen will appear and prompt you to choose between
Normal Mode and Inert Mode. Use the AIR YES or RANGE
SHIFT button to display Inert Mode and press and release POWER
ENTER RESET.
NOTE: To change the Normal Mode oxygen other alarm points, select
Normal Mode and see “Updating the Alarm Point Settings” on
page 116.
10. Press and release POWER ENTER RESET. The alarm point or alarm
operation (oxygen only) will begin to flash.
11. Use AIR YES and RANGE SHIFT to adjust the alarm point to the
desired setting. Keep the following in mind:
The low alarm cannot be set higher than the high alarm and the high
alarm cannot be set lower than the low alarm.
Any alarm setting can be turned off by adjusting it to its lowest
setting. The setting will be displayed as OFF.
12. If you want to continue with the change, press and release POWER
ENTER RESET to accept the setting.
If you want to exit this screen without saving any change to the alarms,
press and release DISPLAY ADJUST NO until you return to the Change
Alarm Point Settings Screen.
MODE SELECT
UP/DOWN THEN ENTER
NORMAL MODE
OXY 0- 40.0 vol%
INERT ALARM
>LO : 5.0 vol%
HI : 10.0 vol%
END

Table of Contents

Other manuals for RKI Instruments EAGLE 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments EAGLE 2 and is the answer not in the manual?

RKI Instruments EAGLE 2 Specifications

General IconGeneral
BrandRKI Instruments
ModelEAGLE 2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals