Eagle Instruction Manual Description 8
Circuit Boards
The Eagle circuit boards analyze, record, control, store, and display
the information collected.
The analog PCB is mounted perpendicular to the base of the
instrument case. It is located in the same half of the case as the sensors
and sample-drawing system. The sensor leads connect to the analog
PCB.
The main PCB is mounted in the top half of the case. It includes the
methane elimination and CAL/SETUP switches.
CAUTION:The circuit boards should be serviced only by authorized repair
personnel.
Methane Elimination Switch
The methane elimination switch (SW1) is mounted near the top right
corner of the main PCB.
For applications where methane is an interfering gas, you can set the
methane elimination switch to eliminate most response to methane
(see “Appendix B: Methane Elimination” on page 52). An external
methane elimination switch is available as an option.
CAL/SETUP Switch
The CAL/SETUP switch (SW2) is mounted near the middle left edge
of the main PCB.
This switch controls the Eagle functions available to the user by
disabling the SHIFT/ button. Without the use of this button, the
user is unable to enter Calibration or Setup mode. (Display mode is
available with either switch setting.) See “Setting User Access” on
page 13 to change the switch setting.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments Eagle Series and is the answer not in the manual?

RKI Instruments Eagle Series Specifications

General IconGeneral
BrandRKI Instruments
ModelEagle Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals