86 • Start Up, Leak Check Mode GX-2012 Operator’s Manual
7. If BP RMNDR is turned on (factory setting is OFF), the screen that
appears next depends on how BP EXPRD is set in the Maintenance
Mode Menu (page 133).
• If the unit is due for a bump test and BP EXPRD is set to
CONFIRM (factory setting), then the following screen displays.
The alarm LED’s and buzzer will pulse several times. After
this, press the RESET SILENCE button until you hear a beep to
continue.
• If the unit is due for a bump test and BP EXPRD is set to NOT
USE, then the following screen displays.
The GX-2012 cannot be used until a bump test has been
performed by selecting BUMP in the Maintenance Mode
menu. See “Performing a Bump Test” on page 111 for bump
testing instructions.
• If a bump test is not due or BP EXPRD is set to NO EFFECT,
then the following screen appears for a few seconds indicating
when the next bump test is due.