Urisys - Operator’s Manual V.
ATTENTION: To remove the printed test report, tear off the paper by pulling it in front across the edge.
() Switch on the Urisys system using the on/off switch at the rear of the instrument.
Self Check
Figure
() The analyzer automatically carries out a self check. The default language
setting for the Urisys system is English. The instrument is factory set to
read Combur
Test
®
UX test strips. It checks that the correct test strip tray
has been inserted. On completion of the self check, the test strip tray is
transported to the start position and the retaining bar opens (see Figure ).
“SELF CHECK OK” is printed together with a date and timestamp.
ATTENTION: If the message E Wrong Tray! is displayed after the Urisys
system is switched on,
press the START button. The selection menu STRIP TYPE is displayed. Select C- by press ing the left
function key.
When the self test is finished a sound is heard from the printer. The display will show Insert Strip. If the paper has already
been inserted, the printout “Self Check OK” is executed with date and time; this happens whenever the instrument is switched
on. If an error occurs, consult the Operator’s Manual or contact your local Roche Diagnostics service representative.
() Upon first use of the analyzer, the message “REPEAT CALIBRATION” appears after a successful self check. Press the
“No” key to attain Ready-to-Measure status. This also happens if the instrument is not used for more than
seven days. To bring the instrument to Ready-to-Measure mode, insert the paper and execute the
calibration.
() To set the language of your choice, press the key sequence “Menu/Setup /Setup /Setup /Language/Other”
(see Sections . and .).
() The Urisys system leaves the factory with the default settings for Printer, Measure and Interface modes, Test Strip
Type, Units, Language, Date/time format (see Section .) and Operator ID. If your laboratory works differently,
individual preferences can be entered via the menus.
() Calibrate the analyzer with Control-Test M (see Section ).
Power-off
We recommend that the analyzer be switched off at the end of each working day and that the mains adapter be unplugged
from the AC wall socket (see Section ., Cleaning).