Urisys  - Operator’s Manual V. 
. Error Messages and Troubleshooting
Error messages are shown in the display but are not printed out. Following  minutes of inactivity, the analyzer switches to
Standby mode. The error message is displayed again when the START button is pressed. In the event that the instrument is
defective, contact your local Roche Diagnostics representative.
E REFERENCE PAD ERROR
MIDDLE!
Cause: The middle portion of the reference pad on the test strip tray is soiled or
damaged.
Action: Switch off the instrument. Carefully clean and dry the pad. Check if it is
damaged (e.g. scratched, etc.). Insert the test strip tray again, then wait for the self
check to finish. If the error message is repeated, replace the reference pad or use
the spare test strip tray. Recalibrate with Control-Test M.
E REFERENCE PAD ERROR
BOTTOM!
Cause: The bottom portion of the reference pad on the test strip tray is soiled or
damaged.
Action: see E
E REFERENCE PAD ERROR TOP! Cause: The top portion of the reference pad on the test strip tray is soiled or
damaged.
Action: see E
E WRONG STRIP !
Cause: The test strip used is different from the one for which the analyzer has
been programmed (Combur

Test
®
UX test strips).
Action: Press the START button. Repeat the measurement with
Combur

Test
®
UX.
E STRIP MEASUREMENT ERROR ! Causes: No test strip is present on the test strip tray, or the strip is incorrectly
positioned on the test strip tray, the urine on the test strip has dried, the test strip
has not been dipped in urine.
Action: Press the START button. Repeat the measurement with a new test strip.
Ensure that all test pads are dipped in the urine sample. Insert the strip correctly
and ensure that the retaining bar is closed properly after START is pressed.
E CALIBRATION ERROR ! Cause: Calibration values differ from those obtained in the last valid calibration.
Action: Press the START button. Repeat the calibration with a new calibration
strip taken from the Control-Test M container. Ensure that the strip is properly
positioned under the clip on the test strip tray (see Section ).
E CALIBRATION INVALID ! Cause: Calibration values are out of tolerance.
Action: Check the reference pad for soiling or damage. Clean if necessary (see
Section .) or use the spare test strip tray. Repeat the calibration with a new
Control-Test M calibration strip (see Section ). If the error message appears
again, the instrument is defective.
E CHIP ERROR ! Cause: The program chip on the right
of the analyzer underneath the printer
cover (see Figure ) is missing, is not
making contact, is defective or contains
an old soft ware version.
Action: Switch off the Urisys 
system. Insert the program chip and
switch the instrument on again. If “CHIP
ERROR” appears again, the instrument
is defective.
Figure 
E MISSING TRAY ! Cause: No test strip tray inserted or test strip tray not inserted far enough to be
engaged by the motor.
Action: Insert the test strip tray correctly (see Section ). Press the START
button.

Other manuals for Roche cobas Urisys 1100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas Urisys 1100 and is the answer not in the manual?

Roche cobas Urisys 1100 Specifications

General IconGeneral
BrandRoche
Modelcobas Urisys 1100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals