Main Applications
Setting up a Run for Purification of Genomic DNA from Whole Blood Samples
41
B
B
Loading the Stage
The graph on the Stage sub-tab is color-coded:
After the resource check, the MagNA Pure 96 Instrument unlocks the load flap. If positions on the Stage
sub-tab are displayed in yellow or red, proceed as follows:
The MagNA Pure 96 Instrument then checks only the rearranged rack.
Once the resource check is successfully completed, the Stage sub-tab of the MagNA Pure 96 Software
looks as follows:
Figure 1 Stage sub-tab after a Successful Resource Check
Proceed with Performing a Run.
Yellow The accessory or disposable is missing.
Red The accessory or disposable is invalid or in the wrong position. For more information
on troubleshooting the relevant messages, see chapter "Troubleshooting" in the
"MagNA Pure 96 Operator’s Guide".
Open the load flap of the MagNA Pure 96 Instrument.
Pull the relevant rack all the way out of the stage.
Rearrange or exchange the relevant components as recommended on the Stage sub-tab.
Insert the rack back into the stage.
Press the rack into the stage until its outline is displayed in blue on the Stage sub-tab.
Close the load flap.