MagNA Pure 96 System, Customer Training Guide V1.0
50
Setting up a Run for Purification of Cellular RNA from Cultured Cells
B
B
Loading the Stage
After the resource check, the MagNA Pure 96 Instrument unlocks the load flap. If positions on the Stage
sub-tab are displayed in yellow or red, proceed as follows:
The MagNA Pure 96 Instrument then checks only the rearranged rack.
Once the resource check is successfully completed, the Stage sub-tab of the MagNA Pure 96 Software
looks as follows:
Figure 2 Stage sub-tab after a Successful Resource Check
Proceed with Performing a Run.
Open the load flap of the MagNA Pure 96 Instrument.
Pull the relevant rack all the way out of the stage.
Rearrange or exchange the relevant components as recommended on the Stage sub-tab.
Insert the rack back into the stage.
Press the rack into the stage until its outline is displayed in blue on the Stage sub-tab.
Close the load flap.