Setting Up the Instrument
R&S
®
RTE
105Getting Started 1326.1026.02 ─ 08
The overall pass/fail result is shown in the "Overall alignment state" field. The
results of the individual alignment steps for each input channel are indicated in
the "Results"
tab. This information is required if problems arise.
7.2 Aligning the Touchscreen
When the device is delivered, the touchscreen is initially calibrated. However, to
ensure that the touchscreen responds to the finger contact correctly, a touch-
screen alignment is required. Only users with administrator rights can fulfill this
task.
Alignment of the touchscreen is useful:
At first use
If the position of the instrument has been changed, and you cannot look
straight on the screen
If another person operates the instrument
If you notice, that touching a specific point on the screen does not achieve the
correct response
1. Press the [SETUP] key.
2. Select the "Screen" tab.
3. Tap "Touchscreen Calibration".
A blinking cross appears in the lower left corner of the screen.
4. Touch and hold the cross until "OK" is shown.
5. Repeat this action for the crosses in the other corners.
6. Tap the R&S logo button in the task bar to display the instrument's user inter-
face.
7.3 Setting the Display Language
You can change the language in which the dialog boxes, result boxes and other
screen information is displayed. A reboot of the instrument is not necessary.
Setting the Display Language