Preparing for Use
R&S
®
RTP
16Getting Started 1337.9946.02 ─ 05
To shut down the instrument
► Press the [Power] key on the front panel.
All current settings are saved, and the software shuts down. The [Power] key
turns orange. The standby power supplies only the power switch circuits.
Now it is safe to power off the instrument.
The "Exit" function in the "File" menu shuts down only the firmware application. To
shut down the instrument completely, also shut down the operating system in the
"Start" menu, or use the [Power] key.
3.3.3 EMI Suppression
Electromagnetic Interference (EMI) may affect the measurement results.
To suppress generated Electromagnetic Interference:
●
Use suitable shielded cables of high quality. For example use double-shielded
RF and LAN cables.
●
Always terminate open cable ends.
●
Note the EMC classification in the data sheet.
3.4 Connecting External Devices
The following interfaces for external devices are provided:
●
USB connectors at the front and rear panel of the instrument
●
Monitor connectors DVI-D and DisplayPort at the rear panel of the instrument
● Connecting USB Devices................................................................................16
● Connecting an External Monitor......................................................................18
3.4.1 Connecting USB Devices
The USB interfaces on the front and rear panels allow you to connect USB devi-
ces directly to the instrument. The number of USB connectors can be increased
by using USB hubs. Due to the large number of available USB devices, there is
almost no limit to the expansions that are possible with the R&S RTP.
Connecting External Devices