Loading...
Improving Level Performance
R&S
®
SMA100B
203User Manual 1178.3834.02 ─ 05
Remote command:
[:SOURce<hw>]:CORRection:VALue? on page 586
UCOR Data
Accesses the standard "File Select" function of the instrument. The provided navigation
possibilities in the dialog are self-explanatory.
Files with user correction values are files with predefined file extension *.uco. When a
file is selected, the dialog indicates the filename.
You can create the file internally in the table editor or externally.
To select an existing file, select "Select List > navigate to the file *.uco > Select"
Use the general editor function to create internally new file or to edit an existing
one.
Use the standard file manager function to load externally created files to the instru-
ment.
Remote command:
[:SOURce]:CORRection:CSET:CATalog? on page 587
[:SOURce<hw>]:CORRection:CSET[:SELect] on page 586
[:SOURce]:CORRection:CSET:DELete on page 588
Edit UCOR Data
Opens the build-in table editor to define a new correction table or edit an existing one.
See also:
Chapter 6.7, "List Editor", on page 188
" Fill... " on page 92
Chapter 7.3.3, "Fill with Sensor", on page 207
7.3.2 List Editor
The "User Correction" and "List Mode" dialogs provide a build-in list editor for defining
the corresponding value pairs.
The table and navigation functions of the UCOR and list mode editors have the same
structure and are therefore summarized below. The access is specified for both modes,
the fields and functions are explained using the example of the list mode.
User Correction

Table of Contents